How you can minimize Risk Factors In Calibration of your Instruments

Choose ISO:IEC 17025:2017 Accredited Calibration Lab service partner. Understand measurement uncertainties and the guard banding process which are being used in the Calibration of your Equipment. Understand the Certificate and Data your Calibration Service Provider is giving you.


What information is included in a Calibration Certificate?

  • Calibration Lab Information
  • Accredited Logo when applicable
  • Details about the Unit Under Test (UUT)
  • Details and Traceability of the equipment used to perform the Calibration
  • The Calibration Method used
  • The results including corrections
  • The uncertainty of measurement
  • The Environmental Conditions which the Calibration was performed
  • Notes about the Calibration

Labelled Calibration Certificate MPC

Requirements of a Calibration Certificate

A. Title
B. Name & Address of the laboratory;
C. Location of the laboratory activities
D. Unique identification that all components are recognized as a portion of a complete report & a clear identification of the end;
E. Name & contact information of the customer
F. Identification of the method used;
G. The condition of the item;
H. Date of receipt of the calibration
I. Date performance of the laboratory
J. Date Issue of the report
K. Statement of effect that results relate only to the item calibrated
L. Authorized Signatories
M. Environmental conditions
N. A statement identifying how the measurements are metrologically traceable
O. Standards Used to Calibrate Equipment
P. Accreditation Body Symbol and Lab Certificate Number (if the calibration service is only covered by the scope of accreditation)
Q. ILAC-MRA Symbol (Optional)


Certificate Data for ISO:IEC 17025-2017 Accredited Calibration

Requirements

R. The measurement uncertainty of the results presented in the same units as that of the measurand or in a term relative to the measurand (%)
S. The results before and after any adjustment or repair, if available
T. Statement of conformity with requirements or specifications

Labelled Calibration Certificate MPC 2

Statements of Pass or Fail Conformance

The uncertainty of measurement has been taken into account when determining compliance with specification.

All measurements and test results guard banded to ensure the probability of false-accept does not exceed 2% in compliance with ANSI/NCSL Z540.3-2006

The status of compliance with the acceptance criteria is reported as:

PASS – Compliant with specification
FAIL – Not compliant with specification
FAIL(z) – The measured value is not within the acceptance limits. However, a portion of the expanded uncertainty of measurement at 95% is within the specified tolerance.
PASS(z) – The measured value is within acceptance limits. However, a portion of the expanded uncertainty of measurement at 95% exceeds the specified tolerance.

Acceptance limits for PFA

The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for a normal distribution corresponds to a coverage probability of approximately 95%, unless otherwise stated.

This calibration report complies with ISO/IEC 17025:2017 and ANSI/NCSL Z540.3-2006, Method 6 — Guard Bands Based on Test Uncertainty Ratio.


Understanding the Scope of Accreditation

ISO/IEC 17025

  • Competence of test and calibration laboratories
  • Demonstrate requirements
    • Able to operate a management system
    • Are technically competent
    • Able to generate technically valid results

ANSI/NCSL Z540.3

  • Technical requirements for calibration of M&TE
  • System for control of the accuracy of M&TE
    • Calibration System performance requirements
    • Calibration laboratory performance requirements
    • Link to use of 17025 accredited Cal labs

Micro Precision Calibration Inc Cert and Scope File

Expanded Uncertainties of Measurement must include all factors that will affect the measurements. Sample factors for an Industrial Scale:

  • Uncertainty of the calibration of standards
  • Possible Drift since the last calibration
  • Resolution of the unit under test

  • Correction of Air Buoyancy
  • Rounding error no load
  • Round error indicated load

  • Repeatability
  • Environmental factors

Calibration Needs. Simplified.

Full stack test equipment and instrument calibration services.

calibration-services

Calibration

Micro Precision is a global leader in equipment calibration. We service a wide range of applications and industries, and we offer on-site and laboratory calibration services.

About Calibration

Instrument Repair Services

Repair

Cost-effective repair services that will get your equipment and operations back in track.

About Repair

Equipment Sales & Rental

Equipment Sales

Find the right equipment to purchase — brand new and used test equipment.

About Equipment Sales

Global Calibration Solutions

Global Solutions

Our global support team offers a range of additional services that enhances your experience. This includes technical support, asset management and more.

About Global Solutions

CDM Software

CDM Software

Managing your equipment calibration status is made easy with our customer data management software.

About CDM Software

Test equipment rental service

Equipment Rental

Flexible short-term solutions for your projects, our leasing service provides a wide database of equipment for rent.

About Equipment Rental


Request A Quote

"*" indicates required fields

This field is for validation purposes and should be left unchanged.
Address*
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
This field is hidden when viewing the form
Accepted file types: jpg, gif, png, pdf, csv, doc, docx, xls, xlsx, Max. file size: 20 MB.

Service Area

If you have any inquiries, recommendations, or simply wish to connect with us, we’d love to hear from you! Please don’t hesitate to reach out to us at

22835 Industrial Place Grass Valley CA 95949

or you can contact us via phone at

(530) 268-1860

or email us at [email protected]

GET DIRECTIONS